Trade-off between spatial details and motion artifact in multi-detector CT: A virtual clinical trial with 4D textured human models

T. Sauer, A. Kapadia, E. Abadi, B. Harrawood, E. Samei, William Paul Segars, Shobhit Sharma

Medical Imaging, 2019

Abstract

In Computed Tomography (CT) imaging, high pitch and wide beam collimations accelerate imaging acquisitions and thus reduce motion artifacts. However, increasing pitch and collimation impact acquisition geometry and thus spatial quality of the images. The purpose of this study was to quantify the effects of pitch and beam collimation on image quality using a realistic virtual clinical trial (VCT) construct. The study used extended-cardio torso (XCAT) phantoms enhanced by synthesizing intra-organ heterogeneities within the lungs and bones. Different amounts of cardiac and respiratory motion were simulated, including heart rates of 0, 60, 90, 120 beats per minute, and respiratory rates of 0, 8, 12, 16 breadths per minute. Each case was imaged using a realistic, scanner-specific, and rapid CT simulator setup based on the geometry and physics of a commercial CT scanner (Siemens Definition Flash), at 120 kV, under multiple pitch values, beam collimations, all at the same dose levels. With the knowledge of the ground truth, the quality of the acquired images was quantified by measuring root mean squared error (RMSE) in the lungs. In general, results indicated that RMSE was higher for the phantoms with more respiratory or cardiac motions. For the pitch experiment, images with higher pitch values had less in-plane motion artifacts. RMSE also increased with increasing the pitch. However, the slope of this trend was found to be a function of motion profile, showing the trade-off…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free