Evidence Hub

Iodine quantification performance with deep silicon-based Photon-Counting CT: A virtual imaging trial study

Panta RK, Yin Z, Grönberg F, Bhattarai M, Abadi E, Segars P, Samei E

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB), 2025

3 citations

Abstract

Purpose: This study investigates the imaging performance of a deep silicon-based photon-counting CT (Si-PCCT) in quantifying iodine contrast through a virtual imaging trial (VIT).

Methods: We developed a VIT framework using Si-PCCT simulator and benchmarked it against a prototype using an ACR phantom for assessing spatial resolution and noise characteristics, and a geometric phantom for iodine quantification. We imaged geometrical phantoms (20 - 40 cm) with iodine concentrations ranging from 1 to 19.7 mg/ml and XCAT human models with iodine contrast at BMI of 19 to 38 kg/m2 across different radiation dose levels (13.9, 27.8, and 41.7 mGy of CTDIvol). We performed material decomposition, reconstructed iodine CT images, and evaluated iodine quantification accuracy.

Results: The Si-PCCT simulator closely matched with the prototype, with differences within 3 % in MTF (f50 and f10) and 3.7 % (fpeak) in NNPS, and Root-Mean-Square Error of 0.12 mg/ml in iodine quantification. The mean absolute errors (MAE) between the estimated and ground-truth iodine concentration were 0.10, 0.25, and 1.80 mg/ml for 20, 30, and 40 cm phantoms, and 0.31, 0.37, and 0.70 mg/ml for XCAT human models with BMIs of 19, 28, and 38 kg/m2, respectively. Similarly, the MAEs were 0.88, 0.45, and 0.31 mg/ml for the geometrical phantoms, and 0.66, 0.5, and 0.46 mg/ml for human models at CTDIvol of 13.9, 27.8, and 41.7 mGy respectively. These results demonstrate accurate iodine quantification performance,…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free