Abstract

BACKGROUND

Cherenkov imaging can be used to visualize the placement of the beam directly on the patient's surface tissue and evaluate the accuracy of treatment planning. However, Cherenkov emission intensity is lower than ambient light. At present, time gating is the only way to realize Cherenkov imaging with ambient light.

PURPOSE

This study proposes preparing a novel carbon quantum dot (cQD) sheeting to adjust the wavelength of Cherenkov emission to obtain the optimal wavelength meeting the sensitive detection region of the camera, meanwhile the total optical signal is also increased. By combining a specific filter, this approach might help in using lower-cost camera systems without intensifier-coupled to accomplish in vivo monitoring of the surface beam profile on patients with ambient light.

METHODS

The cQD sheetings were prepared by spin coating and UV curing with different concentrations. All experiments were performed on the Varian VitalBeam system and optical emission was captured using an electron multiplying charge-coupled device (EMCCD) camera. To quantify the optical characteristics and certify the improvement of light intensity as well as signal-to-noise ratio (SNR) of cQD sheeting, the first part of the study was carried out on solid water with 6 MV and 10 MV photon beams. The second part was carried out on an anthropomorphic phantom to explore the applicability of sheeting when using different radiotherapy materials and the imaging effect of sheeting with the…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.