The Importance of Temporal Resolution for Ultra-High-Resolution Coronary Angiography: Evidence From Photon-Counting Detector CT.
Abstract
PURPOSE: The aim of this study was to assess the effect of temporal resolution on subjective and objective image quality of coronary computed tomography angiography (CCTA) in the ultra-high-resolution (UHR) mode with dual-source photon-counting detector (PCD) CT.
MATERIALS AND METHODS: This retrospective, institutional review board-approved study evaluated 30 patients (9 women; mean age, 80 ± 10 years) undergoing UHR CCTA with a clinical dual-source PCD-CT scanner. Images were acquired with a tube voltage of 120 kV and using a collimation of 120 × 0.2 mm. Gantry rotation time was 0.25 seconds. Each scan was reconstructed using both single-source and dual-source data resulting in an image temporal resolution of 125 milliseconds and 66 milliseconds, respectively. The average heart rate and the heart rate variability were recorded. Images were reconstructed with a slice thickness of 0.2 mm, quantum iterative reconstruction strength level 4, and using the Bv64 and Bv72 kernel for patients without and with coronary stents, respectively. For subjective image quality, 2 experienced readers rated motion artifacts and vessel delineation, or in-stent lumen visualization using 5-point discrete visual scales. For objective image quality, signal-to-noise ratio, contrast-to-noise ratio, stent blooming artifacts, and vessel and stent sharpness were quantified.
RESULTS: Fifteen patients had coronary stents, and 15 patients had no coronary stents. The mean heart rate and heart rate…
How strong is the evidence?
Evidence Hub reads this as a cohort study, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence. Reported size: 30 participants or records.
Read the paper
Related studies
- In silico toxicology protocols
- QM7-X, a comprehensive dataset of quantum-mechanical properties spanning the chemical space of small organic molecules
- Quantum Informed Machine-Learning Potentials for Molecular Dynamics Simulations of CO2's Chemisorption and Diffusion in Mg-MOF-74
- In-silico strategies in nano-drug design: Bridging nanomaterials and pharmacological applications
- mRNA codon optimization with quantum computers
- The computational power of the human brain
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.