Impact of spacing filaments external to hollow fibers on dialysate flow distribution and dialyzer performance
Abstract
A new type of dialyzer (PAN 650 SF Asahi) is analyzed in terms of hydraulic properties, solute clearances and dialysate flow distribution. The new type of dialyzer is a polyacrylonitrile hollow fiber filter, equipped with spacing filaments placed externally to the fibers to facilitate dialysate distribution and avoid channeling. The new filter is compared with a similar filter without spacing filaments. For this purpose, blood and dialysate side clearances have been measured in sequential dialysis session carried out randomly in the same patients. Furthermore, a last generation helical scanner (X-Press/HS1, Toshiba) has been utilized to analyze in vitro the flow distribution of dialysate inside the dialyzer. A contrast medium was injected and a sequence of images has been achieved on a longitudinal section of the dialyzer. This new method permits to avoid any bias due to the cylindrical shape of the dialyzer, since a 10 mm thick rectangular section is analyzed and not the entire body of the filter. The dialyzers equipped with spacing filaments displayed a significant improvement of the dialysate distribution as demonstrated by the radiological pattern. In detail, despite a channeling phenomenon in the peripherical region of the bundle is still present, this is remarkably reduced in comparison with the channelling phenomenon observed in the standard dialyzers. This improved distribution is confirmed by a significant improvement of the solute clearances.
How strong is the evidence?
Evidence Hub reads this as a randomized trial, done in people. Its evidence level is 2 of 5 (High · randomized trial). Randomized trials of a reasonable size are the strongest single-study design.
Read the paper
Related studies
- Preliminary Toxicity Analysis of 3DCRT versus IMRT on the High Dose Arm of the RTOG 0126 Prostate Cancer Trial
- Adaptive Radiotherapy for Head and Neck Cancer: Initial Clinical Outcomes from a Prospective Trial
- American Association of Physicists in Medicine Task Group 263: Standardizing Nomenclatures in Radiation Oncology
- Motion Management Strategies and Technical Issues Associated with Stereotactic Body Radiotherapy of Thoracic and Upper Abdominal Tumors: A Review from NRG Oncology
- High-dose-rate interstitial brachytherapy as a monotherapy for localized prostate cancer: treatment description and preliminary results of a phase I/II clinical trial
- Highly efficient training, refinement, and validation of a knowledge-based plan quality control system for radiotherapy clinical trials
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.