Ultrahigh dose rate pencil beam scanning proton dosimetry using ion chambers and a calorimeter in support of first in‐human FLASH clinical trial

Eunsin Lee, A. Lourenço, J. Speth, N. Lee, A. Subiel, F. Romano, R. Thomas, R. Amos et al.

Medical Physics (Lancaster), 2022

42 citations

Abstract

Abstract Purpose To provide ultrahigh dose rate (UHDR) pencil beam scanning (PBS) proton dosimetry comparison of clinically used plane‐parallel ion chambers, PTW (Physikalisch‐Technische Werkstaetten) Advanced Markus and IBA (Ion Beam Application) PPC05, with a proton graphite calorimeter in a support of first in‐human proton FLASH clinical trial. Methods Absolute dose measurement intercomparison of the plane‐parallel plate ion chambers and the proton graphite calorimeter was performed at 5‐cm water‐equivalent depth using rectangular 250‐MeV single‐layer treatment plans designed for the first in‐human FLASH clinical trial. The dose rate for each field was designed to remain above 60 Gy/s. The ion recombination effects of the plane‐parallel plate ion chambers at various bias voltages were also investigated in the range of dose rates between 5 and 60 Gy/s. Two independent model‐based extrapolation methods were used to calculate the ion recombination correction factors ks to compare with the two‐voltage technique from most widely used clinical protocols. Results The mean measured dose to water with the proton graphite calorimeter across all the predefined fields is 7.702 ± 0.037 Gy. The average ratio over the predefined fields of the PTW Advanced Markus chamber dose to the calorimeter reference dose is 1.002 ± 0.007, whereas the IBA PPC05 chamber shows ∼3% higher reading of 1.033 ± 0.007. The relative differences in the ks values determined from between the linear and quadratic…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free