Effects of storage conditions on the performance of an electrochemical aptamer-based sensor

Chung J, Billante A, Flatebo C, Leung KK, Gerson J, Emmons N, Kippin TE, Sepunaru L et al.

Sensors & diagnostics, 2024

Abstract

The electrochemical aptamer-based (EAB) sensor platform is the only molecular monitoring approach yet reported that is (1) real time and effectively continuous, (2) selective enough to deploy in situ in the living body, and (3) independent of the chemical or enzymatic reactivity of its target, rendering it adaptable to a wide range of analytes. These attributes suggest the EAB platform will prove to be an important tool in both biomedical research and clinical practice. To advance this possibility, here we have explored the stability of EAB sensors upon storage, using retention of the target recognizing aptamer, the sensor's signal gain, and the affinity of the aptamer as our performance metrics. Doing so we find that low-temperature (-20 °C) storage is sufficient to preserve sensor functionality for at least six months without the need for exogenous preservatives.