Long term performance characteristics of an electrochemical nitric oxide analyser
Abstract
BACKGROUND: The NIOX MINO(®) is a nitric oxide (FE(NO)) analyser based on electrochemical technology. It includes a replaceable sensor. Quality control procedures are recommended, but regular calibration is not possible. We aimed to evaluate the performance characteristics of the NIOX MINO(®) to identify if reproducibility changed over time, or with different sensors. Also, there are reports that reproducibility of FE(NO) may be reduced in patients with high FE(NO): our secondary aim was to address this issue.
METHODS: Reproducibility in 24 separate sensor-analyser units was calculated on three occasions over two months in 17 patients. These included 9 patients whose FE(NO) was high (mean 80 ppb) and 8 in whom FE(NO) was low (mean 16 ppb).
RESULTS: One device failed quality control testing. For the remaining 23 sensor-analyser combinations, the mean coefficient of variation was 4.0% (range 1.2-7.2%) at baseline, 3.6% (range 2.0-7.0) at one month, and 3.6% (range 1.6-7.6%) at two months. The 95% C.I. for the mean limits of agreement for FE(NO) was ± 4.2 ppb (range 0.9-9.6 ppb), ± 3.8 ppb (range 1.6-6.9 ppb) and ± 3.2 ppb (range 1.2-6.8 ppb) respectively (NS). The limits of agreement exceeded the manufacturer's specifications (± 5 ppb) in 0 devices at baseline, 3 (13%) at one month, and 5 (22%) at two months.
CONCLUSIONS: Reproducibility of FE(NO) using the NIOX MINO(®) was within clinically acceptable limits (± 10 ppb) and was generally stable. However, with time, a…
How strong is the evidence?
Evidence Hub reads this as a randomized trial, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence. Reported size: 17 participants or records.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.