Abstract
Abstract In our study, a commercially available continuous ketone monitoring device captured β-Hydroxybutyrate (BHB) dynamics during exogenous ketosis but revealed a gradual decline day-to-day BHB concentrations over 14 days in both ketone ester and placebo groups, likely reflecting sensor drift.