Evaluation of a novel continuous glucose measurement device in patients with diabetes mellitus across the glycemic range
Abstract
BACKGROUND: This glucose clamp study assessed the performance of an electrochemical continuous glucose monitoring (CGM) system for monitoring levels of interstitial glucose. This novel system does not require use of a trocar or needle for sensor insertion.
METHOD: Continuous glucose monitoring sensors were inserted subcutaneously into the abdominal tissue of 14 adults with type 1 or type 2 diabetes. Subjects underwent an automated glucose clamp procedure with four consecutive post-steady-state glucose plateau periods (40 min each): (a) hypoglycemic (50 mg/dl), (b) hyperglycemic (250 mg/dl), (c) second hypoglycemic (50 mg/dl), and (d) euglycemic (90 mg/dl). Plasma glucose results obtained with YSI glucose analyzers were used for sensor calibration. Accuracy was assessed retrospectively for plateau periods and transition states, when glucose levels were changing rapidly (approximately 2 mg/dl/min).
RESULTS: Mean absolute percent difference (APD) was lowest during hypoglycemic plateaus (11.68%, 14.15%) and the euglycemic-to-hypoglycemic transition (14.21%). Mean APD during the hyperglycemic plateau was 17.11%; mean APDs were 18.12% and 19.25% during the hypoglycemic-to-hyperglycemic and hyperglycemic-to-hypoglycemic transitions, respectively. Parkes (consensus) error grid analysis (EGA) and rate EGA of the plateaus and transition periods, respectively, yielded 86.8% and 68.6% accurate results (zone A) and 12.1% and 20.0% benign errors (zone B). Continuous EGA yielded 88.5%,…
How strong is the evidence?
Evidence Hub reads this as a clinical trial, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence. Reported size: 14 participants or records.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.