Evaluating surface EMG control of motorized wheelchairs for amyotrophic lateral sclerosis patients
Abstract
Background: This study evaluated a novel control method for patients unable to independently control powered wheelchairs. Patients with amyotrophic lateral sclerosis often require a wheelchair but struggle with sufficient hand dexterity required for joystick control making them a population that needs this type of control method.
Methods: The study employed a novel control mechanism, using electromyography surface sensors applied to temporalis muscles able to measure the myoelectric voltage. Pattern and magnitude control of muscle contraction allowed for steering intention recognition and were used to manipulate their power wheelchair joystick. Four patients ages 51 to 69, two female and two male with amyotrophic lateral sclerosis, conducted Wheelchair Skills Test developed by Dalhousie University and were surveyed on the experience's Clinical Global Impression of Change.
Results: Findings showed independent steering was capable for patients without hand function and provided recommendations for improved human-machine interface. All patients demonstrated the ability to engage the system, with varying precision, for driving their wheelchair in a controlled environment.
Conclusions: Three patients in the pilot trial reported the highest score of clinical global impression of change, all of whom had lost independent control of their wheelchair joystick. Patient four retained impaired hand dexterity for joystick control and reported negative impression of change, comparatively.…
How strong is the evidence?
Evidence Hub reads this as a clinical trial, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.