Evidence Hub

Resistance to Acetaminophen Interference in a Novel Continuous Glucose Monitoring System

Calhoun P, Johnson TK, Hughes J, Price D, Balo AK

Journal of diabetes science and technology, 2018

66 participants · 49 citations

Abstract

Acetaminophen (APAP) can cause erroneously high readings in real-time continuous glucose monitoring (rtCGM) systems. APAP-associated bias in an investigational rtCGM system (G6) was evaluated by taking the difference in glucose measurements between rtCGM and YSI from 1 hour before to 6 hours after a 1-g oral APAP dose in 66 subjects with type 1 or type 2 diabetes. The interference effect was defined as the average post-dose (30-90 minutes) bias minus the average baseline bias for each subject. The clinically meaningful interference effect was defined as 10 mg/dL. The G6 system's overall mean (±SD) interference effect was 3.1 ± 4.8 mg/dL (one-sided upper 95% CI = 4.1 mg/dL), significantly lower than 10 mg/dL. The G6 system's resistance to APAP interference should provide reassurance to those using the drug.

How strong is the evidence?

Evidence Hub reads this as a clinical trial, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence. Reported size: 66 participants or records.

How the five levels of evidence work. The level is assigned automatically from the abstract and can be wrong, so check the paper.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free