Continuous Glucose Monitor Interference With Commonly Prescribed Medications: A Pilot Study
Abstract
Background: Reliability of continuous glucose monitors (CGM) is a prerequisite for therapeutic dosing of insulin without the need for confirmatory blood glucose meter measurements. Interference of CGMs with commonly prescribed substances has not been extensively evaluated.
Methods: We sought to undertake a novel pilot study to determine the susceptibility of FDA-approved CGM systems (Medtronic Guardian Sof-Sensor, Dexcom G4 Platinum) to erroneous readings in the presence of common medications. CGMs were placed on the abdomen of healthy subjects 48 hours prior to study. Subjects were admitted to the Clinical Research Trials Unit (CRTU) on the evening before study and fed a standard supper. The following morning, an oral medication was administered in the fasted state and blood was sampled for 9 hours. CGM values were compared to ambient glucose (measured with YSI) to observe variations in CGM readings. Microdialysis catheters were also placed in the abdomen to sample interstitial fluid (ISF) for drug concentrations.
Results: Nineteen healthy drug-naïve subjects without diabetes participated in the study. A drug/substance was tested up to a maximum of nine times on separate occasions. Comparison of CGM glucose patterns to actual plasma glucose concentrations show several drugs, including lisinopril, albuterol, and acetaminophen, appear to interfere with commonly used CGM devices. Wine also interfered with CGM readings.
Conclusions: We conclude there is some evidence of CGM…
How strong is the evidence?
Evidence Hub reads this as a clinical trial, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.