Comparison of Force Distribution during Laryngoscopy with the C-MAC D-BLADE and Macintosh-Style Blades: A Randomised Controlled Clinical Trial
Abstract
Background: The geometry of a laryngoscope’s blade determines the forces acting on the pharyngeal structures to a relevant degree. Knowledge about the force distribution along the blade may prospectively allow for the development of less traumatic blades. Therefore, we examined the forces along the blades experienced during laryngoscopy with the C-MAC D-BLADE and blades of the Macintosh style. We hypothesised that lower peak forces are applied to the patient’s pharyngeal tissue during videolaryngoscopy with a C-MAC D-BLADE compared to videolaryngoscopy with a C-MAC Macintosh-style blade and direct laryngoscopy with a Macintosh-style blade. Beyond that, we assumed that the distribution of forces along the blade differs depending on the respective blade’s geometry. Methods: After ethical approval, videolaryngoscopy with the D-BLADE or the Macintosh blade, or direct laryngoscopy with the Macintosh blade (all KARL STORZ, Tuttlingen, Germany), was performed on 164 randomly assigned patients. Forces were measured at six positions along each blade and compared with regard to mean force, peak force and spatial distribution. Furthermore, the duration of the laryngoscopy was measured. Results: Mean forces (all p 0.128). With the D-BLADE, the forces were highest at the blade’s tip. In contrast, the forces were more evenly distributed along the Macintosh blades. Videolaryngoscopy took the longest with the D-BLADE (p = 0.007). Conclusions: Laryngoscopy with the D-BLADE resulted in…
How strong is the evidence?
Evidence Hub reads this as a randomized trial, done in people. Its evidence level is 2 of 5 (High · randomized trial). Randomized trials of a reasonable size are the strongest single-study design. Reported size: 164 participants or records.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.