Comparison of two devices for measuring exercise transcutaneous oxygen pressures in patients with claudication
Abstract
BACKGROUND: We aimed at estimating the agreement between the Medicap" (photo-optical) and Radiometer• (electrochemical) sensors during exercise transcutaneous oxygen pressure ( tcp02) tests. Our hypothesis was that although absolute starting values (tcp02rest: mean over 2 minutes) might)e different, tcpOTchanges over time and the minimal value of the decrease from rest of oxygen pressure (DROPmin) results at exercise shall be concordant between the two systems. taneously, one of each system on the chest, on each buttock and on each calf.
RESULTS: Seventeen Medicap" probes disconnected during the tests. tcp02rest and DROPmin values were higher with Medicap• than with Radiometer•, by 13.7 ± 17.1mmHg and 3.4 ± 11.7 mmHg, respectively. Despite the differences in absolute starting values, changes over time were similar between the two systems. The concordance between the two systems was approximately 70% for classification of test results from DROPmin.
CONCLUSIONS: Photo-optical sensors are promising alternatives to electro-chemical sensors for exercise oximetry, provided that miniaturization and weight reduction of the new sensors are possible.
PATIENTS AND METHODS: Forty seven patients with arterial claudication ( 65 ± 7 years) performed a treadmill test with 5 probes each of the electro-chemical and photo-optical devices simul-
How strong is the evidence?
Evidence Hub reads this as a cohort study, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.