The clinical benefit of cardiac resynchronization therapy optimization using a device-based hemodynamic sensor in a patient with dilated cardiomyopathy: a case report
Abstract
IntroductionResults on the evolution of the clinical status of patients undergoing cardiac resynchronization therapy with a defibrillator after automatic optimization of their cardiac resynchronization therapy are scarce. We observed a rapid and important change in the clinical status of our non-responding patient following activation of a sensor capable of weekly atrioventricular and interventricular delays' optimization.Case presentationA 78-year-old Caucasian man presented with dilated cardiomyopathy, left bundle branch block, a left ventricular ejection fraction of 35 %, New York Heart Association class III/IV heart failure, and paroxysmal atrial fibrillation. Our patient was implanted with a cardiac resynchronization device with a defibrillator and the SonRtip atrial lead. Right ventricular and left ventricular leads were also implanted. Because of the recurrence of atrial fibrillation, the automatic optimization was set off at discharge. Consequently, the device did not optimize atrioventricular and interventricular delays (programming at discharge: 125 ms for the atrioventricular delay and 0 ms for the interventriculardelay). Our patient was treated with an anti-arrhythmic drug. Five months after implantation, his clinical status remained impaired (left ventricular ejection fraction = 30 %). The SonR signal amplitude had also decreased from 0.52 g to 0.29 g. Nevertheless, because our patient was no longer presenting with atrial fibrillation, the anti-arrhythmic…
How strong is the evidence?
Evidence Hub reads this as a case report, done in people. Its evidence level is 5 of 5 (Very low · case / narrative / animal / in vitro / in silico). Case reports, narrative reviews, and animal, cell or computer-model studies give the weakest evidence for human health.
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.