Accuracy of an electrochemical sensor for measuring capillary blood ketones by fingerstick samples during metabolic deterioration after continuous subcutaneous insulin infusion interruption in type 1 diabetic patients
Abstract
OBJECTIVE: This study was designed to test the accuracy of capillary ketonemia for diagnosis of ketosis after interruption of insulin infusion. RESEARCH DESIGN AND METHODS: A total of 18 patients with type 1 diabetes treated by external pump were studied during pump stop for 5 h. Plasma and capillary ketonemia and ketonuria were determined every hour from 7:00 A.M. (time 0 min = T0) to 12:00 P.M. (time 300 min = T300). Plasma beta-hydroxybutyrate (beta-OHB) levels were measured by an enzymatic end point spectrophotometric method, and capillary beta-OHB levels were measured by an electrochemical method (MediSense Optium meter). Ketonuria was measured by a semiquantitative test (Ketodiastix). Positive ketosis was defined by a value of >/=0.5 mmol/l for ketonemia and >/=4 mmol/l (moderate) for ketonuria. RESULTS: After stopping the pump, concentrations of beta-OHB in both plasma and capillary blood increased significantly at time 60 min (T60) compared with T0 (P /=250 mg/dl, plasma and capillary ketonemia were found to be more frequently positive (85 and 78%, respectively) than ketonuria (59%) (P = 0.017). The time delay to diagnosis of ketosis was significantly higher for ketonuria than for plasma ketonemia (212 +/- 67 vs. 140 +/- 54 min, P = 0.0023), whereas no difference was noted between plasma and capillary ketonemia. CONCLUSIONS: The frequency of screening for ketosis and the efficiency of detection of ketosis definitely may be improved by the use of capillary blood…
Read the paper
Related studies
- Nano Biosensors: Properties, applications and electrochemical techniques
- Electrochemically Synthesized MIP Sensors: Applications in Healthcare Diagnostics
- Toward a Harmonization for Using in situ Nutrient Sensors in the Marine Environment
- Electrochemical determination of rutin by using NiFe2O4 nanoparticles-loaded reduced graphene oxide
- Application of Polyoxometalate-based composites for sensor systems: A review
- Sensitive Electrochemical and Thermal Detection of Human Noroviruses Using Molecularly Imprinted Polymer Nanoparticles Generated against a Viral Target
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.