Evidence Hub

Mitigating X-ray-Induced Damage in CD-SAXS Metrology of EUV Resists

Wieser PA, Wu X, Tiwale N, Fukuto M, Freychet G, Gann E, Goldfarb D, Lavoie C et al.

ACS applied materials & interfaces, 2026

Abstract

Accurate characterization of nanoscale features in photoresist patterns is critical for advancing extreme ultraviolet lithography. Critical-dimension small-angle X-ray scattering (CD-SAXS) enables reconstruction of nanoscale patterns from angle-dependent scattering data, but measurements of organic, chemically active materials, such as photoresists, are strongly limited by beam-induced damage. Here, we demonstrate that CD-SAXS can be successfully applied to organic resists when X-ray exposure is carefully managed to mitigate beam damage. We quantify beam-induced damage in multiple resist systems by monitoring diffraction pattern evolution under continuous X-ray exposure, establishing a fluence threshold of ∼1000 photons/nm2 at 16.1 keV for the resist systems and measurement conditions investigated. We evaluate complementary strategies to extend the usable measurement window: reversing the sample orientation to exploit beam attenuation by the substrate (about sixfold dose reduction), increasing the photon energy, and reducing the angular sampling density.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free