Evidence Hub

Dual X-Ray and Optical System for In Situ Monitoring of Solids Fraction in Tailings

Sraw TK, Yu B, Liu X, Ng J, Sedgwick A, Gupta M, Fedosejevs R, Tsui YY

Sensors (Basel, Switzerland), 2026

Abstract

A sensing apparatus was developed to quantify solids fraction within a 3 m test column, demonstrating real-time monitoring of tailings sedimentation. This integrated system employs a low-power infrared laser coupled with scattering detectors alongside a low-activity radioactive source and detector. Both methodologies utilize non-destructive principles to evaluate the concentration of solids. Specifically, X-ray transmission through the medium provides a measure of the solids fraction, as X-ray attenuation corresponds to the solids percentage. Concurrently, the optical component detects radiation backscatter from the sample. To ensure precision, both instruments underwent calibration using reference materials. Experimental evaluations were conducted using two configurations: a fixed array and a depth-profiling setup. The former utilized a 2.5 m plastic column with integrated sensors placed within a settling vessel to acquire data at discrete depths. Conversely, the depth-profiling configuration involved a submersible mobile unit lowered into the column to characterize the solids profile at arbitrary depths.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free