Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T

Lee C, Xu J, Yang B, Noeske R, Magnotta VA

2026

127 participants

Abstract

Subject motion and gradient heating-induced frequency and phase offsets result in spectral misalignment during single-voxel MR spectroscopy (SVS) acquisitions. Several methods have been presented to align the spectra, but their performance on a 7T system is unclear. This study aimed to evaluate the practical importance of four retrospective correction methods, namely creatine fitting, residual water, spectral registration, and cross-correlation. SVS data were collected from 127 participants (88/39 female/male, 39±15 years) using a semi-localization by adiabatic selective refocusing sequence at 7T. Changes in the spectral linewidth, signal-to-noise ratio (SNR), similarity (mean similarity matrix (SI mean )), and metabolite quantification (concentration, relative Cramer-Rao lower bounds (rCRLB)) after correction were evaluated. A Wilcoxon signed rank test was used to evaluate these changes. The p -values were adjusted using the false discovery rate ( q 0.60). The correction only showed a weak to moderate effect of the changes in the metabolite quantification (effect size<0.60), suggesting that whether the correction results in more reliable metabolite quantification requires further validation.