Measurement of photon beam backscatter from collimators to the beam monitor chamber using target-current-pulse-counting and telescope techniques
Abstract
Backscattered radiation (BSR) arising from field-defining collimators and entering the beam monitor chamber (BMC) may contribute to observed variations in medical linear accelerator photon beam output with collimator setting. Measuring the magnitude of such contributions for particular accelerators under specified operating conditions is therefore important when attempting to understand and model accelerator head scatter. The present work was conducted to confirm some backscatter measurements for collimating jaws reported previously and to extend these to include other accelerators and a multileaf collimator (MLC). BSR reaching the BMC from the jaws of Clinac 600C, 2100C and 2300CD accelerators and from an MLC on the 2300CD was investigated using both target-current-pulse-counting and telescope methods. Our measurements show that for the Clinac 600C BSR-dependent output variations are negligible. However, for the 2100C and 2300CD BSR-dependent relative output increased in an almost linear fashion, by up to 2.4% for 15 and 18 MV beams, and by up to 1.7% for 6 MV beams, as the field size varied from 5 x 5 cm2 to 40 x 40 cm2. The magnitude of BSR dependent upon collimator location in the head, as expected, thereby contributing to the collimator exchange effect. An earlier study at our centre using the telescope method had reported higher BSR levels. This discrepancy was resolved when corrections for telescope block and room scatter, previously assumed negligible, were made.
Read the paper
Related studies
- Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
- Fast Monte Carlo based joint iterative reconstruction for simultaneous 99mTc/ 123I SPECT imaging
- Dopamine Transporter SPECT with 12-Minute Scan Duration Using Multiple-Pinhole Collimators
- Dual modality intravascular catheter system combining pulse-sampling fluorescence lifetime imaging and polarization-sensitive optical coherence tomography
- Use of capillary optics as a beam intensifier for a Compton x-ray source
- Comparison between Two Adaptive Optics Methods for Imaging of Individual Retinal Pigmented Epithelial Cells
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.