Monitor chamber backscatter for intensity modulated radiation therapy using multileaf collimators
Abstract
Backscattered radiation into the machine monitor chamber can affect the machine output variation, with changes in field size and shape. For intensity modulated radiation therapy (IMRT) where many field, which may have small dimensions, are summed to give an intensity modulated field, the magnitude of backscatter will be different due to both the backscattering surface area changing, and the delivered monitor units being larger than for the equivalent static field. The effect of backscatter variation with field size for a Philips SL15 accelerator has been investigated at 8 MV for static and IMRT fields both in the standard clinical operating condition where an anti-backscatter plate is fitted, and also for a case where the anti-backscatter plate has been removed. The results show that in the absence of the anti-backscatter plate the variation in output between a 4 cm by 4 cm field and a 40 cm by 40 cm field size due to backscattered radiation was 5% for static fields. The anti-backscatter plate reduced this variation to less than 1%. When the accelerator operated in IMRT mode, with the backscatter plate in place, changes in the output due to additional backscattered radiation were less than 0.3%. With the backscatter plate removed, the outputs were lower, indicating the presence of additional backscattered radiation. It can be concluded that for the Philips MLC and SL accelerator with its anti-backscatter plate, the effects of backscattered radiation can be ignored for both…
Read the paper
Related studies
- Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
- Fast Monte Carlo based joint iterative reconstruction for simultaneous 99mTc/ 123I SPECT imaging
- Dopamine Transporter SPECT with 12-Minute Scan Duration Using Multiple-Pinhole Collimators
- Dual modality intravascular catheter system combining pulse-sampling fluorescence lifetime imaging and polarization-sensitive optical coherence tomography
- Use of capillary optics as a beam intensifier for a Compton x-ray source
- Comparison between Two Adaptive Optics Methods for Imaging of Individual Retinal Pigmented Epithelial Cells
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.