An analytic model of pinhole aperture penetration for 3D pinhole SPECT image reconstruction
Abstract
Photons penetrate the attenuating material close to the aperture of pinhole collimators in nuclear medicine, broadening the tails of point spread functions (PSFs) and degrading the resolution of planar and SPECT images. An analytic approximation has been developed that models this penetration contribution to the PSF for knife-edge point pinhole apertures. The approximation has the form exp(-gamma r), where r is the distance on the detector surface from the projection of the point source through the pinhole. The rolloff coefficient gamma is a function of the photon energy, point source location and the design parameters of the collimator. There was excellent agreement between measured values of gamma from photon transport simulations of I-131 point sources (364 keV emission only) and theoretical predictions from the analytic formula. Predicted gamma values from the analytic formula averaged 25% greater than measured values from experimental I-131 point source acquisitions. Photon transport simulations were performed that modelled the 364 keV and less abundant 637 and 723 keV emissions and scatter within the scintillation crystal. Measured gamma values from these simulations averaged 12% greater than the experimental values, indicating that about half of the error between the analytic formula and the experimental measurements was due to unmodelled 637 and 723 keV emissions. The remaining error may be due in part to scatter in the pinhole region and backscatter from gamma…
Read the paper
Related studies
- Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
- Fast Monte Carlo based joint iterative reconstruction for simultaneous 99mTc/ 123I SPECT imaging
- Dopamine Transporter SPECT with 12-Minute Scan Duration Using Multiple-Pinhole Collimators
- Dual modality intravascular catheter system combining pulse-sampling fluorescence lifetime imaging and polarization-sensitive optical coherence tomography
- Use of capillary optics as a beam intensifier for a Compton x-ray source
- Comparison between Two Adaptive Optics Methods for Imaging of Individual Retinal Pigmented Epithelial Cells
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.