Computational models of direct and indirect x-ray breast imaging detectors for in silico trials

Aunnasha Sengupta, A. Badal, A. Makeev, A. Badano

Medical Physics (Lancaster), 2022

5 citations

Abstract

BACKGROUND

To facilitate in silico studies that investigate digital mammography (DM) and breast tomosynthesis (DBT), models replicating the variety in imaging performance of the DM and DBT systems, observed across manufacturers are needed.

PURPOSE

The main purpose of this work is to develop generic physics models for direct and indirect detector technology used in commercially available systems, with the goal of making them available open source to manufacturers to further tweak and develop the exact in silico replicas of their systems.

METHODS

We recently reported on an in silico version of the SIEMENS Mammomat Inspiration DM/DBT system using an open-source GPU-accelerated Monte Carlo x-ray imaging simulation code, MC-GPU. We build on the previous version of the MC-GPU codes to mimic the imaging performances of two other FDA-approved DM/ DBT systems, such as Hologic Selenia Dimensions (HSD) and the General Electric Senographe Pristina (GSP) systems. In this work, we developed a hybrid technique to model the optical spread and signal crosstalk observed in the GSP and HSD systems. MC simulations are used to track each x-ray photon till its first interaction within the x-ray detector. On the other hand, the signal spread in the x-ray detectors is modeled using previously developed analytical equations. This approach allows us to preserve the modeling accuracy offered by MC methods in the patient body, while speeding up secondary carrier transport (either electron-hole pairs or…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free